Channel Master Operation Manual
P/N 95B-6001-00 (February 2006) page 81
10.8 Sensor Commands
The following commands let you set the sensor functions.
>S?
Available Commands:
ST C -------------------- Set temperature units (0=c; 1=f; 2=k)
SZ ---------------------- Zero Pressure Sensor
S? ---------------------- Display S-Command Menu
10.8.1 Sensor Command Descriptions
This section lists the sensor commands.
SZ – Zero Pressure Sensor
Purpose Zeros the pressure sensor.
Format SZ
Recommended Setting. Use as needed.
Description This command zeros the pressure sensor at the specific loca-
tion where the Channel Master will be used.
ST – Temperature Units
Purpose Sets the units for the temperature sensor in the PC2 test.
Format STn
Range n = 0 (Celsius), 1 (Fahrenheit), 2 (Kelvin)
Default ST0
Description Sets the units for the temperature sensor as output by the PC2
command (see “PC2 – Display Sensor Data,” page 76).
NOTE. This command does not change the temperature units in the
Variable Leader Data (see Figure 28, page 106).